Tuesday, November 25, 2008

"NEW" DALE601 Electrical Safety Analyzer



Detailed Description

NEW UNIT comes with one-year manufacture warranty against defect and workmanship
Description
The DALE601 Electrical Safety Analyzer is the industry's lowest priced full function Electrical Safety Analyzer
Unique to the DALE601 is its capability to make a broad range of external measurements. A clearly labeled single master function switch guides the user through a complete measurement procedure

The DALE601 measures Line Voltage, Instrument Current, Ground Resistance, Earth and Chassis Leakage Current, Patient Lead-to-Ground, Lead-to-Lead and Lead Isolation (M.A.P.) Current, External Gradient, Probe and Transducer Isolation Current and Point-to-Point Resistance.

The device also tests all 120-volt standard wall outlets for the proper wiring. The green and an amber light indicate five normal or fault conditions

Heavy-duty switches change polarity, open and close the neutral and select Chassis or Earth Leakage Current measurements. A four-wire Kelvin bridge eliminates reading errors in resistance due to cable length and contact resistance. True RMS measurements are provided for all current readings. The AAMI test load is utilized.

The DALE601 has five (5) universal EKG lead connectors that allow for any type of lead connection.There is also a single slide switch that gives the user a choice

Features

· Small, lightweight, self-contained portable instrument
· Master range meter delivers accurate measurment for each test
· Accurate test measurements with 4-wire Kelvin bridge
· Test method complies with AAMI load per ANSI/AAMI ES1-1993 or IEC601-1
· Leakage Current can be measured in Open/Closed Neutral, Normal/Reversed Polarity and Open/Closed Ground.
· All Leakage Current measurements are displayed directly in �Amps.
· Broad range of external measurements · Five (5) Universal EKG Lead Connectors
· The DALE601 is designed for 120 V operation

Specifications
Mains Voltage
117 Volts, 50/60 Hz

Instrument Current
Range: 0 - 19.99 Amps.
Resolution: 10 mA
Accuracy: 5 % R
Frequency:
DC & 25 Hz - 1kHz: 1.0% R
1KHz - 100 KHz: 2.5% R
100 KHz - 1 MHz: 5% R
Controls
1. Function Switch - 9 position rotary
2. Lead Switch - 6 position rotary
3. Neutral Switch - 2 position rocker (Open/Closed)
3. Neutral Switch - 2 position rocker (Open/Closed)
4. Polarity Switch - 3 position rocker (Normal/Off/Reversed)
5. Leakage Switch - 2 position rocker (Chassis/Earth)

DUT Power
117 VAC, 16 Amps@ 120 V, 10 Amps @ 230 V

Physical Size:
4.0 x 7.8 x 1.6 inches (102 x 196 x 41 mm)